(一) Error display
According to | error | note |
---|---|---|
neutral electrode red light | neutral electrode link failure | Detection of neutral electrode is good |
(二) Audio output table
serial number | Audio properties | Character description | note |
---|---|---|---|
1 | 1 normal buzzer | Bass for 0.5 seconds | Normal initialization |
2 | Fault sound | Low high-pitched tune up two: drip drip | neutral electrode failure |
3 | Confirm sound | High and low tune | neutral electrode failure recovery |
4 | Radiofrequency ablation output Effective buzzer | high continuous tone | Effective output during radiofrequency ablation |
5 | Radiofrequency condensate output Effective song | Bass continuant | Effective output during radiofrequency coagulation |
6 | Fault sound | same tuning two tone: drip drip | Indicates a fault, see error display |
(三) Common problems to deal with
phenomenon | why | Elimination method |
---|---|---|
Open the power switch, the equipment without any response | 1,the net power outage; 2,connect the power cord is not; 3,fuse fusing. | 1,connect the power supply; 2,connect the power cord; 3,replace the fuse. |
In normal work setting, significantly lower output power intensity Settings or equipment can't work normally | 1,neutral electrode poor contact; 2,using method is not correct | 1,the correct connection good neutral electrode; 2,repair or replace accessories; 3,please industry to solve |